2018-01-01



The Setaflash Series 8 'ActiveCool' is an automated closed cup flash point tester with a temperature range of 10 to 130°C (50 to 266°F) with no external cooling required. When connected to a suitable chilled water supply, the range is extended down to -20°C (-4°F).

There are two modes of operation; Flash / No Flash and Ramp. The ramp mode increases the sample temperature at a rate of 2°C/min and automatically dips the ignitor every 1°C up to 100°C (every 2°C above 100°C), until either a flash is detected or the end of test temperature is reached.

ADDITIONAL INFORMATION

The automatic flash detection system is tuned to ignore halo and other false flash effects, enhancing precision and accuracy. Barometric pressure is measured by an internal barometer, and the correction is automatically applied to the test result.

Peltier cell technology is used for heating and cooling, ensuring precise temperature control and rapid cooldown after a test is completed. The ignitor uses a new electrical hot wire (a gas version is available - 82110-0). The bright colour LCD display shows test parameters and the flash point result. The text and colour of the top line indicates instrument status. All functions are accessed via menus, navigated by the keypad. Up to 64 test results can be stored in memory, each tagged with a sample reference and operator name. Results can be downloaded to a PC or printer via the RS232C port (USB or LAN ports options available). The firmware is upgradeable for future developments.

SUPPLIED WITH: Ignitor, 2ml syringe, mains lead and instruction manual.

SPECIFICATIONS

Description

Temperature Range

Air: 10 to 130°C (50 to 266°F) Water Assisted: -20 to 130°C (-4 to 266°F) using 99300-2

Sample Size

2 or 4ml according to method

Ignitor

Electrical Hot Wire

Test duration, Rapid Equilibrium Mode

1 minute below 100°C, 2 minutes above 100°C, FAME 1 minute or user defined 1 to 99 minutes

Test duration, Ramp Mode

Typically 7 minutes

Cool Down Time

80 to 20°C (176 to 68°F) in 4 minutes

Sample Cup Material

Corrosion resisting steel insert, aluminium cup

Heating / Cooling Method

Peltier Cell

Computer Interface

RS232C (USB & LAN options)

Voltage

100 to 250V, 50/60Hz (2x AA dry cell batteries required for internal clock)

Power

300W (maximum)

Size (HxWxD)

30 x 34 x 38cm

Weight

8kg

CCCN Code:

Tariff 90268020

FEATURES

10°C to 130°C (-20°C with external water supply)

Small sample size, 2 or 4ml

Flash/No Flash & Ramp modes

Electric ignitor

Automatic dipping and flash detection

ActiveCool electronic Peltier cooling

64 Test memory & RS232 interface

°C or °F temperature display

Corrosion resisting insert

METHODS

ASTM D3278, ASTM D3828, ASTM D7236, ASTM E502, BS 3900 Part A13, BS 6664 Parts 3 & 4, CHIPS Regulations, EPA 1020 A & B, IP 523, IP 524, IP 534, ISO 3679, ISO 3680, UN Class 3 Non-viscous Flammable Liquids

DETAILS OF METHODS

Method

Description

CHIPS Regulations

Chemicals Hazard Information and Packaging for Supply

EPA 1020 A & B

Setaflash Closed-Cup Method for determining ignitability

ASTM D3828

Standard Test Methods for Flash Point by Small Scale Closed Tester

ASTM D3278

Standard Test Methods for Flash Point of Liquids by Small Scale Closed-Cup Apparatus - The test methods are applicable to paints, enamels, lacquers, varnishes, and related products having a flash point between 0 and 110°C (32 and 230°F) and Viscosity lower than 150 St at 25°C (77°F).

ISO 3680

Methods of test for paints. Tests on liquid paints (excluding chemical tests). Test for flash/no flash (rapid equilibrium method)

UN Class 3 Non-viscous Flammable Liquids

Small scale flash point testers are universally specified in many international regulations to define and classify "flammable" and "combustible" substances.

ASTM E502

Standard Test Method for Selection and Use of ASTM Standards for the Determination of Flash Point of Chemicals by Closed Cup Methods

ISO 3679

Determination of flash point. Rapid equilibrium closed cup method

BS 6664 Parts 3 & 4

Methods of test for paints. Tests on liquid paints (excluding chemical tests). Test for flash/no flash (closed cup equilibrium method)

BS 3900 Part A13

Test for flash/no flash (rapid equilibrium method)

IP 523

Determination of flash point. Rapid equilibrium closed cup method

IP 524

Methods of test for paints. Tests on liquid paints (excluding chemical tests). Test for flash/no flash (rapid equilibrium method)

ASTM D7236

Flash Point by Small Scale Closed Cup Ramp Method

IP 534

Flash Point by Small Scale Closed Cup Ramp Method

£0.00