2018-02-06

Kyoritsu Indonesia - Earth Testers
KEW 4105DL





4105DL [Cable Reel Set Model] NEW
4105DL-H [Hard Case Model] NEW
• 3pole and 2pole Earth Resistance measurement (0.01Ω-2000Ω)
• Waterproof design (IP67)
• Rotary Switch makes the user interface very intuitive
• Large LCD Display with Backlight
• Led to monitor correct / non correct auxiliary earth spike resistance
• Earth Voltage Measurement (AC/DC 0-300V)
• CAT Ⅳ 100V



Cable Reel Set Model

Hard Case Model

Earth resistance
measurement

20Ω

200Ω

2000Ω

Measuring range

0.00 - 2000 Ω

Display range

0.00 - 20.99 Ω

0.0 - 209.9 Ω

0 - 2099 Ω

Accuracy*1

±1.5%rdg±0.08 Ω*2

±1.5%rdg±4dgt

Auxiliary earth resistance*3

<10kΩ

<50kΩ

<100kΩ

Comparator reference value

10Ω

100Ω

500Ω

Earth voltage measurement

Measuring range

0 - 300 V AC (45 - 65Hz)

±0 - ±300 V DC

Display range

0.0 - 314.9 V

0.0 - ±314.9 V

Accuracy

±1%rdg±4dgt

Overload protection

Earth resistance : 360V AC (10 seconds)

Earth voltage : 360V AC (10 seconds)

Applicable Standards

IEC 61010-1, CAT IV 100 V /CAT III 150 V /CAT II 300 V Pollution degree 3

IEC 61010-2-030,IEC 61010-031, IEC 61557-1, -5

IEC 60529 IP67, IEC 61326-1, -2-2, EN50581(RoHS)

Power source

LR6(AA)(1.5V) × 6

Dimensions

121(L) × 188(W) × 59(H) mm (including case lid)

Weight

Approx. 690g (including batteries and case lid)

Accessories
for 4105DL

7127B(Simplified measurement probe)

8041(Auxiliary earth spikes[2 spikes/1set])

9121(Shoulder strap)

7267(Cable reel for Earth resistance tester (red))

7268(Cable reel for Earth resistance tester (yellow))

7271(Earth resistance test leads)

9190(Carrying case)

LR6(AA) × 6, Instruction manual

Accessories
for 4105DL-H

7127B(Simplified measurement probe)

8041(Auxiliary earth spikes[2 spikes/1set])

9121(Shoulder strap)

7266(Earth resistance test leads[red-20m, yellow-10m, green-5m/1set])

9191(Hard case)

LR6(AA) × 6, Instruction manual

Optional

7272(Precision measurement Cord set)

8259(Adapter for measurement terminal)

*1 For precision measurement, auxiliary earth resistance should be 100 Ω ±5% or less.
*2 At simplified measurement add ±0.10 Ω to the specified accuracy.
*3 Accuracy within the auxiliary earth resistance: ±5% rdg ±10 dgt.

Accessories/Optional Accessories /Related Products

MODEL 7266

MODEL 7127B

MODEL 8041

MODEL 9121

MODEL 9190

MODEL 9191

</li

Show more